General

Detection modes UV-Vis absorbance
Fluorescence intensity
Luminescence
Fluorescence polarization
Time-resolved fluorescence
Alpha
Read methods Endpoint, kinetic, spectral scanning, well area scanning
Microplate types 6- to 1536-well plates
Other labware supported Petri and cell culture dishes
Take3 Micro-Volume Plates
Temperature control 4-Zone incubation to 65 °C with Condensation Control™
+0.2 °C at 37 °C
Shaking Linear, orbital, double orbital
Software Gen5™ Microplate Reader and Imager Software
Gen5 Secure for 21 CFR Part 11 compliance (option)
Automation BioStack and 3rd party automation compatible
BioSpa 8 Automated Incubator compatible
CO2 and O2 control (option) Range: 0 - 20% (CO2); 1 - 19% (O2)
Control Resolution: + 0.1% (CO2 and O2)
Stability: +0.2% at 5% CO2; +0.2%at 1% O2
Models for CO2 /O2 or CO2 only are available
Barcode reader 1D and 2D camera-based scanner

Absorbance

Light source Xenon flash
Detector photodiode
Wavelength selection monochromator
Wavelength range 230 - 999 nm, 1 nm increments
Monochromator bandwidth <3 nm (230 - 285 nm), <6 nm (>285 nm)
Dynamic range 0 - 4.0 OD
Resolution 0.0001 OD
Pathlength correction yes
Monochromator wavelength accuracy +2 nm
Monochromator wavelength repeatability +0.2 nm
OD accuracy <1% at 2.0 OD
<3% at 3.0 OD
OD linearity <1% from 0 to 3.0 OD
OD repeatability <0.5% at 2.0 OD
Stray light 0.03% at 230 nm
Reading speed (kinetic) 96 well: 6 seconds

Fluorescence Intensity

Light source Xenon flash
Detector Dual top PMTs
Single top PMT (option)
Low noise PMT (bottom filter system)
Red shifted PMT (top/bottom monochromator system)
Wavelength selection Quad monochromators (top/bottom)
Filters (top/bottom)
Wavelength range Monochromators: 250 - 850 nm
Filters (dual PMT): 200 - 850 nm
Monochromator bandwidth Variable, from 3 to 50 nm in 1 nm increments
Dynamic range 7 decades
Sensitivity Filters:
Fluorescein 0.2 pM (4 amol/well, 20 µL, 384-well low volume plate) - top
Fluorescein 1 pM (10 amol/well, 10 µL, 1536-well plate) - top
Fluorescein 1 pM (0.1 fmol/well, 100 µL, 384-well plate) - bottom

Quad monochromator:
Fluorescein 2 pM (40 amol/well, 20 µL, 384-well low volume plate) - top
Fluorescein 2.5 pM (0.25 fmol/well, 100 µL, 384-well plate) - bottom
Reading speed (kinetic) 96 well: 6 seconds
384 well: 11 seconds
1536 well: 25 seconds

Luminescence

Wavelength range 300 - 700 nm
Dynamic range > 6 decades
Sensitivity 5 amol ATP typical (384-well low volume plate)

Fluorescence Polarization

Light source Xenon flash
Detector Dual PMT or single PMT (option)
Wavelength selection Filters
Wavelength range 280 - 850 nm
Sensitivity 1 mP standard deviation at 1 nM fluorescein

Time-Resolved Fluorescence

Light source Xenon flash or TRF laser (option)
Detector Dual PMT or single PMT (option)
Wavelength selection Filters (top/bottom)
Wavelength range Monos: 250 - 850 nm
Filters (dual PMT): 200 - 850 nm
Sensitivity With TRF laser: 5 fM (384-well low volume plate)
With Xenon flash lamp: 40 fM (384-well low volume plate)

Alpha Detection

Light source 100 mW 680 nm laser
Detector PMT
Wavelength selection Filters (top)
Sensitivity 100 amol LCK peptide (384-well low volume plate)
Read speed 96 well: 30 seconds
384 well: 1 minute 50 seconds
1536 well: 7 minutes 20 seconds

Reagent Dispensers

Supported detection modes All modes
Number 2 syringe pumps
Supported labware 6- to 384-well plates, Petri dishes
Dead volume 1.1 mL with back flush
Dispense volume 5 - 1000 µL in 1 µL increments
Dispense accuracy + 1 µL or 2%
Dispense precision < 2% at 50-200 µL

Physical Characteristics

Power 250 Watts maximum consumption.
Dimensions Without TRF laser:
15.4" W x 20.7" D x 16.1" H
(39.2 x 52.5 x 40.8 cm)

With TRF laser:
15.4" W x 24.2" D x 16.1" H
(39.2 x 61.4 x 40.8 cm)
Weight Without TRF laser: 78 lbs. (35.4 kg)
With TRF laser: 100 lbs. (45.3 kg)

Regulatory

Regulatory CE and TUV marked. RoHS Compliant. Models for In Vitro Diagnostic use are available.