Synergy 2 Multi-Mode Microplate Reader

Detection > Filter-based Multi-Mode Microplate Readers

Synergy 2 Multi-Mode Microplate Reader

Overview:

Read 2µL samples!
The Synergy 2 is BioTek's high-performance multi-mode plate reader designed for the life science research and drug discovery markets. Its fluorescence detection system uses deep blocking filters and dichroic mirrors for the best level of performance in research and screening applications. The dedicated absorbance detection system is monochromator-based, providing the best level of flexibility in this read mode. And, by utilizing dedicated optical elements for each individual detection technique, Synergy 2 offers uncompromised performance and the lowest limits of detection.

Detection modes include Fluorescence Intensity, Fluorescence Polarization, Time-Resolved Fluorescence, AlphaScreen, Luminescence and UV-visible Absorbance. Its three broad-spectrum light sources have been chosen for optimal illumination and excitation in all applications.

The Synergy 2 is controlled by BioTek's Gen5™ Reader Control and Data Analysis Software, is certified for HTRF®, DLR™ and LanthaScreen™, and is compatible with AlphaScreen assays.

US Federal Government customers:

This product is available on GSA contract. Contact BioTek Customer Care with questions or submit a Quote Request Form to receive pricing and information on specific models (please note GSA Contract in the Comments field).

HTRF® and the HTRF logo are registered trademarks of Cisbio.
Dual-Luciferase is a trademark of Promega Corporation and is registered with the U.S. Patent and Trademark office.
AlphaScreen® is a registered trademark of Perkin Elmer.
Transcreener® is a registered trademark of BellBrook Labs.
The LanthaScreen™ and GeneBLAzer® technologies from Invitrogen have been validated on BioTek instruments.


Features:


Models:

Synergy 2 is a modular instrument. Read modes and options are available as individual modules at the time of purchase, allowing you to make the best use of your budget. In addition, existing units can be upgraded later when your research requirements expand.

 Synergy™ 2 Custom Configurations
 Contact BioTek for part #

 Module

 Description

 S  Synergy 2 base model.  Includes temperature control to 65ºC, Shaking, and Gen5™.
 L  Luminescence
 F  Fluorescence Intensity Top and Bottom.  Includes Tungsten-Halogen light source.
 P  Fluorescence Polarization.  Requires F module.
 T  Time-Resolved Fluorescence.  Requires F module.  Includes high energy DPR Xenon Flash Lamp.
 A  UV-Visible Absorbance (monochromator-based).  Includes SQ Xenon Flash Lamp.
 D  Dual Reagent Dispenser

Red shifted PMT (900nm) option available instead of standard low noise PMT (700 nm).
3 mm and 1.5 mm diameter bottom fiber options available instead of standard 5 mm bottom fiber.


Specifications:

General
Detection method Fluorescence, Time Resolved Fluorescence, Fluorescence Polarization, Alpha Screen/ AlphaLisa, Luminescence, UV-Visible Absorbance
Read method End-point, kinetic, spectral scanning, well-area scanning
Microplate types 6- to 1536-well plates (luminescence 1-384)
PCR plates
Compatible with Take3 Multi-Volume Plate with 2 µL microspots
Temperature control 4°C above ambient to 65°C
Shaking Yes
Software Gen5™ Included
Control through USB or serial port
Automation Yes
AlphaScreen/AlphaLISA
Light source Tungsten Halogen Lamp
Sensitivity 100 amol of biotinylated-LCK-P peptide, 25 µL/well in 384-well plate
Dynamic range >6 decades
Reading speed 2 minutes typical for 96-well plate
Detection system Ultra Low Noise PMT
Absorbance
Light source SQ Xenon Flash
Wavelength selection Monochromator
Wavelength range 200 - 999 nm, 1 nm increment
Bandpass 2.4 nm
Dynamic range 0 - 4.0 OD
Resolution 0.0001 OD
Pathlength correction Yes
Monochromator wavelength accuracy +/- 2 nm
Monochromator wavelength repeatability +/- 0.2 nm
OD accuracy < 1% at 2 OD typical
< 3% at 3.0 OD typical
OD linearity < 1% from 0 to 3.0 OD typical
OD repeatability < 0.5% at 2.0 OD typical
Stray light 0.03% at 230 nm typical
Reading speed 96: 11 seconds
384: 22 seconds
1536: 42 seconds
Fluorescence Intensity
Sensitivity Fluorescein 1 pM typical (0.2 fmol/well 96-well plate; 0.1 fmol/well 384-well plate) - Top
Fluorescein 5 pM typical (1 fmol/well 96-well plate; 0.5 fmol/well 384-well plate) - Bottom
Light source Tungsten Halogen High Energy DPR Xenon Flash (optional)
Wavelength selection Deep blocking bandpass filters/dichroic mirrors
Wavelength range 300 - 700 nm with tungsten lamp (900 nm option)
200 - 700 nm with xenon lamp (900 nm option)
Dynamic range > 6 decades
Bandpass Filter dependent
Detection system PMT
Reading speed 96: 11 seconds
384: 22 seconds
1536: 42 seconds
Luminescence
Sensitivity 10 amol ATP typical (flash)
Wavelength range 300 - 700 nm
Dynamic range > 6 decades
Detection system Low noise PMT
Fluorescence Polarization
Light source Tungsten Halogen High Energy DPR Xenon Flash (optional)
Sensitivity 3 mP at 1 nM fluorescein typical
Wavelength selection Deep blocking filters/mirrors
Wavelength range 400 - 700 nm
Time-Resolved Fluorescence
Light source High Energy DPR Xenon Flash
Sensitivity Europium 60 fM typical (12 amol/well 96-well plate; 6 amol/well 384-well plate)
Wavelength selection Deep blocking filters/mirrors
Wavelength range 200 - 700 nM (900 nm option)
Reagent Dispensers
Number 2 syringe pumps
Dispense volume 5 - 1000 µL in 1 µL increment
Dead volume 1.1 mL, 100 µL with backflush
Physical Characteristics
Power 250 Watts max.
Dimensions 17"W x 17.5"D x 14.5"H (43.5 x 44.5 x 37.3 cm)
Weight 60 lbs (27 kg)
Regulatory
Regulatory For in vitro diagnostic use. All BioTek microplate instrumentation is CE and ETL marked. For details on approvals and standards compliance, please contact BioTek.