Synergy H4 Hybrid Multi-Mode Microplate Reader

Detection > Hybrid Multi-Mode Microplate Readers

Synergy H4 Hybrid Multi-Mode Microplate Reader

Overview:

Read 2µL samples!

Synergy™ H4 with Hybrid Technology™ is a patent-pending multi-mode detector that combines the optical systems of Synergy Mx and Synergy 2 in one compact instrument. It is the ideal instrument for research and drug discovery applications when having to choose between flexibility and performance is not an option. The monochromator-based optics provides a high level of flexibility; any wavelength can be used from the low UV to the near infrared. The quadruple grating optical system, equipped with a variable bandpass selection system, is ideal for spectral scanning applications. The filter-based optics use dichroic mirrors for enhanced performance. This system is faster and more sensitive than the monochromator optics. A dual reagent dispenser option is available for inject and read assays such as flash luminescent assays and fluorescent ion channel assays. Additional read modes include: fluorescence polarization, time resolved fluorescence and AlphaScreen/AlphaLISA.

Watch this interview at SelectScience.tv with Marina Bruss from Analytica 2010 to learn more about the Synergy H4 Hybrid Multi-Mode Microplate Reader.

HTRF® and the HTRF logo are registered trademarks of Cisbio.
Transcreener
® is a registered trademark of BellBrook Labs.
The LanthaScreen™ and GeneBLAzer® technologies from Invitrogen have been validated on BioTek instruments.


Features:

Take3 Multi-Volume Plate


Models:

Synergy™ H4 is BioTek's second generation Hybrid reader. Like Synergy 4, it offers the combined benefits of filter-based and monochromator-based detection.

 Synergy™ H4
Part # Absorbance LUM FI (Filters & Monos) FP TRF Injectors Software
H4MF           Gen5™ included
H4MFD        
H4MFA        
H4MFAD      
H4MLF        
H4MLFD      
H4MLFA      
H4MLFAD    
H4MLFPTA  
H4MLFPTAD
H4MFPT      
H4MFPTD    
H4MLFPT    
H4MLFPTD  
H4MFPTA    
H4MFPTAD  

Specifications:

General
Wavelength Selection Hybrid Technology™ (patent-pending) Quadruple Monochromators and Filters / Dichroics
Detection method Fluorescence, Time-Resolved Fluorescence, Fluorescence Polarization, AlphaScreen/AlphaLISA®, Luminescence, UV-Visible Absorbance
Read method End-point, kinetic, spectral scanning, well-area scanning
Microplate types Monochromator system 1- to 384-well plates
Filter system 1- to 1536-well plates
PCR plates
Compatible with Take3™ Multi-Volume Plate with 2 µL microspots
Temperature control 4° C above ambient to 65° C ± 0.5° C at 37° C
Shaking Yes
Software Gen5™ Included Control through USB or serial port
Automation Yes
AlphaScreen/AlphaLISA
Light source Tungsten Halogen Lamp
Sensitivity 100 amol of biotinylated-LCK-P peptide, 25 µ­L/well in 384-well plate
Dynamic range >6 decades
Reading speed 2 minutes typical for 96-well plate
Detection system Ultra Low Noise PMT
Absorbance
Light source Xenon Flash Lamp
Wavelength selection Monochromator
Wavelength range 230 - 999 nm, 1 nm increment
Bandpass 2 nm (230-285 nm), 4 nm (>285 nm)
Dynamic range 0 - 4.0 OD
Resolution 0.0001 OD
Pathlength correction Yes
Monochromator wavelength accuracy +/- 2 nm
Monochromator wavelength repeatability +/- 0.2 nm
OD accuracy < 1% at 2.0 OD typical
< 3% at 3.0 OD typical
OD linearity < 1% from 0 to 3.0 OD typical
OD repeatability < 0.5% at 2.0 OD typical
Stray light 0.03% at 230 nm typical
Reading speed 96: 11 seconds
384: 22 seconds
Fluorescence Intensity
Sensitivity Monochromators:
Fluorescein 2 pM typical (0.2 fmol/well 384-well plate) -Top
Fluorescein 2.5 pM typical (0.25 fmol/well 384-well plate) - Bottom
Filters/mirrors:
Fluorescein 1 pM typical (0.1 fmol/well 384-well plate)
Light source Tungsten Halogen Lamp
High Energy Xenon Flash Lamp
Wavelength selection Double grating monochromators (Top/Bottom) and,
Deep blocking filters / dichroic mirrors (Top)
Wavelength range Monochromators: 250 - 900 nm
Filters: 200 - 700 nm (900 nm option)
Dynamic range Monochromators: 5 decades
Filters/mirrors: > 6 decades
Bandpass Monochromators: variable 9, 13.5, 17, 20 nm
Filters: filter-dependant, from 5 nm to >100 nm
Detection system Red shifted PMT for monochromator system
Low Noise PMT for filter-based system
Reading speed 96: 11 seconds
384: 22 seconds
1536: 42 seconds
Luminescence
Sensitivity 10 amol ATP typical (flash)
Wavelength range 300 - 700 nm
Dynamic range > 6 decades
Fluorescence Polarization
Light source Tungsten Halogen
Sensitivity 3 mP at 1 nM fluorescein typical
Wavelength selection Deep blocking filters / dichroic mirrors (Top)
Wavelength range 400 - 700 nm (300-900 nm option)
Time-Resolved Fluorescence
Light source High Energy Xenon Flash Lamp
Sensitivity Europium 60 fM typical with filters (6 amol/well in 384-well plate)
Wavelength selection Deep blocking filters / dichroic mirrors (Top)
Double grating monochromator (Top / Bottom)
Wavelength range Filters: 200 - 700 nm (900 nm option)
Monochromators: 250 - 900 nm
Reagent Dispensers
Number 2 syringe pumps
Dispense volume 5 - 1000 µL in 1 µL increment
Dead volume 1.1 ml, 100 µL with back flush
Physical Characteristics
Power 250 Watts max.
Dimensions 17"W x 20.9"D x 15"H (43.5 x 53.1 x 38.1 cm)
Weight 78 lbs (35 kg)
Regulatory
Regulatory All BioTek microplate instrumentation is CE and ETL marked. For details on approvals and standards compliance, please contact BioTek.