ELx405 Microplate Washer
Over the last 20 years, BioTek has become the leader in microplate washing. This proven expertise in the design and manufacture of the highest quality microplate washers has evolved into the ELx405™ Microplate Washer. Precise, fast and easy to use -- the ELx405 includes a variety of models for washing standard and deep well plates, each with its own innovative features.
To meet the ever growing demand for automated washing of microsphere based assays, the industry leading ELx405 Microplate Washer is now available with modules for both vacuum filtration and biomagnetic separation. BioTek's latest ELx405 enhancements offer full plate washing of polystyrene and magnetic microspheres used in a growing number of multiplex assays and bead-based ELISAs, such as those developed on the Luminex® xMAP® technology platform. An available vacuum filtration feature automates the wash steps of filter bottom plates with four adjustable vacuum levels for optical performance with filter pore sizes ranging from 0.45 µm to 1.2 µm. Developed with leaders in genotyping, gene expression and protein assays, the ELx405's biomagnetic separation incorporates high-energy neodymium iron boron magnets for rapid separation of micrometer and nanometer beads with superior retention. The ELx405 Washer is Luminex® xMAP® approved.
The ELx405 HT is designed specifically to rapidly wash 384-well microplates and includes BioTek's patented Dual-Action™ manifold to ensure optimal performance and easy maintenance.
The ELx405 Select is the market leader in 96-/384-well washing. The Select also has our patented Dual-Action™ manifold, providing independent control of both dispense and aspirate manifolds for precise overfill washing and overflow protection. This unique design washes both 96- and 384-well plates without the need for any hardware changes. With its angled dispense tubes, full control of dispense and aspiration rates and unique X, Y, Z positioning, the ELx405 Select is ideal for the most demanding wash requirements. The ELx405 Select CW is specially designed to accommodate loosely-adherent cell lines. The new ELx405 Select Deep Well is the world's first deep well microplate washer that aspirates and dispenses into 96- and 384-well plates <50 mm in height. ANSI/SBS standard microplates are also accommodated resulting in a total solution for all of a laboratory's solid bottom microplate washing needs.
For those applications that use only 96-well microplates, the standard ELx405 Microplate Washer is an ideal choice for performance and value! The ELx405 is the most precise and reliable washer available. Many unique features lower background absorbance values and increase the signal-to-noise ratio of typical ELISA assays.
The ELx405's patent pending Ultrasonic Advantage™ offers effortless washer manifold maintenance. Scientists regard unexpected clogging of manifold tubes to be the largest contributor to assay failure. Cleaning these tubes is a time consuming and labor intensive process. BioTek's Ultrasonic Advantage™ incorporates the power of ultrasonic cleaning to make the ELx405 the first self-maintaining microplate washer. BioTek's Liquid Handling Control Software provides full PC control of the ELx405 in a familiar Microsoft Windows® environment, while its intuitive StepWise™ protocol creation makes LHC Software easy to use even for novice users.
A complete solution, the ELx405 offers 96- and 384-well plate washing for multiplex bead assays, ELISAs and cellular assays along with filtration-to-waste processes... all together in a single, robot-compatible design.
US Federal Government customers:This product is available on GSA contract. Contact BioTek Customer Care with questions or submit a Quote Request Form to receive pricing and information on specific models (please note 'GSA Contract' in the Comments field).
xMAP® is a registered trademark of Luminex® Corporation.
Hardware:
- Accommodates 96- and 384- standard and deep well microplates
- Polystyrene and magnetic bead assays along with standard ELISAs
- Patented Dual-Action™ manifold on the ELx405™ HT and ELx405™ Select provides independent control of the aspirate and dispense manifolds for precise overfill washing and overflow protection
- Internal positive displacement pump eliminates the use of troublesome pressure delivery systems and allows the use of any size dispense reservoir
- Built-in priming trough allows automatic manifold priming and rinsing during both operation and maintenance modes and eliminates the need for blank plates when used in automated robotic systems
- Fully variable dispense volumes, flow, and aspiration rates provide total control of the washing cycles from a gentle cell-washing application to vigorous wash for high-binding assays
- Built-in fluid flow, fluid detection, and vacuum sensing provide complete protection for unattended operation
- Optional external valve module connects four reagent bottles with automatic switching between reagents
- BioStack compatible for automated washing of up to 50 ANSI/SBS standard height microplates
Patented Dual-Action™ manifold design for 96- and 384-well plates.
The ELx405™ Select Deep Well Washer aspirates and dispenses into full size 96- and 384-well blocks with unattended operation.
Software:
- 75 programs
- Link up to 10 programs (if valve module is attached, links multiple buffers automatically)
- Pre/AutoPrime
- 10 cycles/wash
- Crosswise aspiration
- Fully programmable:
- dispense height
- aspiration height
- aspiration delay
- soak
- shake
- aspiration rate
- horizontal aspirate and dispense position
- External programming/downloading available through USB or RS-232 ports
Model |
Part # |
96-/384- well |
384-well |
96-well |
Deep Well |
Cell Washing |
Buffer Switching |
Ultrasonic |
Biomagnetic Separation |
Vacuum Filtration |
| ELX405™ HT | ELX405HT | • |
||||||||
| ELX405HTM | • |
• |
||||||||
| ELX405HTMF | • |
• |
• |
|||||||
| ELX405HTVMF | • |
• |
• |
• |
||||||
| ELX405HTS | • |
• |
||||||||
| ELX405HTV | • |
• |
||||||||
| ELX405HTVS | • |
• |
• |
|||||||
| ELX405HT2 | • |
|||||||||
| ELX405HT2S | • |
• |
||||||||
| ELX405HTV2 | • |
• |
||||||||
| ELX405HTV2S | • |
• |
• |
|||||||
| ELX405HTV2SM | • |
• |
• |
• |
||||||
| ELx405™ Select | ELX405U | • |
||||||||
| ELX405UD | • |
• |
• |
|||||||
| ELX405UM | • |
• |
||||||||
| ELX405UMF | • |
• |
• |
|||||||
| ELX405UV | • |
• |
||||||||
| ELX405UVD | • |
• |
• |
• |
||||||
| ELX405UVMF | • |
• |
• |
• |
||||||
| ELX405US | • |
• |
||||||||
| ELX405USD | • |
• |
• |
• |
||||||
| ELX405UVS | • |
• |
• |
|||||||
| ELX405UVSD | • |
• |
• |
• |
||||||
| ELX405UCW | • |
• |
||||||||
| ELX405UCWD | • |
• |
• |
• |
||||||
| ELX405UCWS | • |
• |
• |
|||||||
| ELX405UCWSD | • |
• |
• |
• |
• |
|||||
| ELX405UCWSMF | • |
• |
• |
• |
• |
|||||
| ELX405UCWV | • |
• |
• |
|||||||
| ELX405UCWVD | • |
• |
• |
• |
• |
|||||
| ELX405UCWVS | • |
• |
• |
• |
||||||
| ELX405UCWVSD | • |
• |
• |
• |
• |
• |
||||
| ELX405UCWVSM | • |
• |
• |
• |
• |
|||||
| ELx405™ | ELX405UCWVSMF | • |
• |
• |
• | • | • | |||
| ELX405R | •
|
|||||||||
| ELX405RM | • |
• | ||||||||
| ELX405RMF | • |
• | • | |||||||
| ELX405RS | • |
• |
||||||||
| ELX405RSM | • |
• | • | |||||||
| ELX405RSMF | • |
• |
• |
• |
||||||
| ELX405VR | • |
• |
||||||||
| ELX405VRMF | • |
• |
• | • | ||||||
| ELX405VRS | • |
• |
• |
|||||||
| ELX405VRSM | • |
• |
• |
• |
||||||
| ELX405VRSMF | • |
• |
• | • | • |
| ELx405™ 96-/384-well Microplate Washer Configurations |
||
| Feature | Design | |
| HT | Faster 384-well microplate washing |
|
| M | Biomagnetic Separation (Magnetic beads) | |
| F | Vacuum Filtration (Polystyrene Beads) | |
| S | Ultrasonic Advantage™ |
|
| V | Automatic 4-Buffer Switching | |
| 2 | 96-/384-well microplate washing |
|
| U | 96-/384-well microplate washing |
|
| D | Deep well microplates |
|
| CW | Optimized cell washing for loosely adherent monolayers | |
| R | 96-well microplate washing |
|
| General | |
| Assays | Magnetic bead, polystyrene bead (model dependent)
ELISA Cell-based assays (model dependent) Filtration-to-waste processes (model dependent) |
| Separation | Biomagnetic separation, vacuum filtration (model dependent) |
| Microplate Types | 96- and 384-well (model dependent) Deep well - up to 50 mm (model dependent) Low profile and standard height Solid and filter bottom (model dependent)
|
| Shaking | Programmable in minutes and seconds, up to 60 minutes Intensities – slow, medium, fast or variable |
| Soak Time | Programmable in minutes and seconds, up to 60 minutes |
| Magnet | High strength 96- and 384-well designs Choice of 2 designs, custom for BioTek microplate washers: Flat - Flat-bottom well - beads pulled to band across well bottom - Round-bottom well - beads pulled to button at well bottom Ring - beads pulled to 4-zone ring at well bottom |
| Safety/Convenience/Maintenance | Waste and vacuum sensing, fluid and flow detection Aerosol cover Adjustment utility for plate positioning Overflow protection Pre-programmed maintenance routines |
| Automation | BioStack™ Microplate Stacker - up to 50 ANSI/SBS standard height plates (optional)
|
| Onboard Software | 2 x 24 character backlit LCD display 25 alphanumeric soft keys Create, edit or run multiple protocols |
| Software | Liquid Handling Control™, for PC wash protocol programming and execution (optional) |
| Operating Temperature | 15°C to 30°C (59°F to 86°F) |
| Humidity | 10% to 85%, non-condensing |
| Electrical | 100 - 240 Volts AC 50/60 Hz, <8.0 A |
| Washing | |
| Manifold Types | 96-well washing: 96-tube (8 X 12) manifold 96-/384-well washing: Dual-Action™ 96-tube (8 X 12) manifold 384-well washing: Dual-Action™ 192-tube (16 x 12) manifold |
| Washing Speed | Solid bottom plates: 3 asp. / disp. cycles, 300 µL/well, 96 wells, 96-tube manifold, final aspirate: <30 seconds 3 asp. / disp. cycles, 100 µL/well, 384 wells, 96-tube manifold, final aspirate: <80 seconds 1 asp. / disp. cycle, 400 µL/well, 384 wells, 192-tube manifold, final aspirate: <20 seconds Filter bottom plates: Variable, based on wash parameters |
| Fluid Delivery | One positive displacement pump |
| Vacuum Filtration | Selectable vacuum levels (final at 30 seconds): 0.45 µM 96-well plates: Lowest: -111 mmHg Low: -200 mmHg Medium: -405 mmHg High: -561 mmHg 0.45 µM 384-well plates: Lowest: -108 mmHg Low: -196 mmHg Medium: -371 mmHg High: -563 mmHg 1.2 µM 96-well plates: Lowest: -117 mmHg Low: -203 mmHg Medium: -409 mmHg High: -550 mmHg 1.2 µM 384-well plates: Lowest: -117 mmHg Low: -198 mmHg Medium: -387 mmHg High: -450 mmHg Vacuum filtration time range: 5 to 999 seconds |
| Volume Range | 25 - 3000 µL/well (model dependent) Selectable in 1 µL increments |
| Buffer Selection | Automatic switching for up to 4 wash buffers (model dependent) |
| Flow Rates | 1 - 11 Includes low flow, cell wash rates (model dependent) |
| Wash Cycles | 1 - 10 |
| Dispense Precision | ≤3% CV (model dependent) |
| Residual Volume | Solid bottom plates ≤2 µL/well Filter bottom plates 96-well: Average increase weight of plate ≤1.2 g after dispensing 300 µL to 0.45 µm plate, filtration 30 seconds, low vacuum, blotted 384-well: Average increase weight of plate ≤4.0 g after dispensing 80 µL to 1.2 µm plate, filtration 10 seconds, low vacuum, blotted |
| Sterilization | Chemical:
70% ethyl or isopropyl alcohol and 0.5% sodium hypochlorite solution |
| Ultrasonic Cleaning | Built-in ultrasonic bath for protein and salt crystal buildup Programmable in hours and minutes, up to 60 minutes (model dependent) |
| Supply Bottle Volume | 4 L or 10 L (optional) |
| Physical Characteristics | |
| Dimensions | Depth: 17" (43.2 cm) Width: 14" (35.6 cm) Height: 10" (25.4 cm) - 405, 405 Select, 405 HT 11" (27.9 cm) - 405 Select Deep Well |
| Weight | 30 lbs (13.5 kg) |
| Regulatory | |
| Regulatory | For In Vitro Diagnostic use. All BioTek microplate instrumentation is CE and ETL marked |
Specifications are subject to change.
Optional

BioStack Microplate Stacker

Liquid Handling Control Software
Standard (Included) | ||
Miscellaneous | ||
| 7102209 | Aerosol Cover |
|
Operators Manual | ||
| MANUAL | Operator's manuals can be obtained via the Customer Resource Center | |